Model | LS-65A/LS-65B /LS-65C | LS-90A/LS-90B/LS-90C | LS-115A/LS-115B/LS-115C |
Capacity | 65L | 90L | 115L |
Dimensions (L×W×H) | 664×660×994.5 mm | 664×660×994.5 mm | 664×660×1200 mm |
Sterilization Chamber Dimensions | φ400×530 mm | φ400×730 mm | φ400×920 mm |
Rated Power(W) | 2800W | 3400W | 4500W |
Stainless Steel Baskets | φ370×200×2 | φ370×200&φ370×400 | φ370×400×2 |
Working Voltage | AC 220±10%50Hz |
Temperature Display Accuracy | 0.1℃ |
Operating Environment | 5℃~40℃、relative humidity10%~85% |
Sterilization Chamber Material | SUS304 stainless steel |
Sterilization Working Temperature | 105℃~143℃ |
Sterilization Time Pre-set Range | 1min~9999mins |
Melting Temperature Pre-set Range | 60℃~100℃ |
Melting Time Pre-set Range | 1min~9999mins |
Insulation Temperature Pre-set Range | 45℃~60℃ |
Insulation Time Pre-set Range | 1min~9999mins |
Door Opening Temperature Pre-set Range | The solid,agar,and waste modes are from 40℃to 99℃,while the liquid mode is from 40℃to 80℃ |
Clock Function Timer Pre-set Range | The operating time can be set(year,month,day,hour,minute),and it allows querying of the current time,total sterilization cycles,as well as fault occurrence times and error codes. |
Pressure Gauge Display Range | 0~0.6Mpa |
Pressure | The chamber design pressure is 0.35 MPa.Each unit comes with a corresponding product quality certificate and an original certificate issued by the special equipment manufacturing supervision and inspection department. |
Pressure Safety Protection | Two independent pipelines for measurement and control,with dual mechanical and electronic pressure protection. |
Exhaust Controland Water Tank | Fully automatic internal exhaust(exhaust temperature settings are detailed in the manual),with a built-in large-capacity steam condensate collection bottle and water tank. |
Cover Opening Method and Interlocking Device | Unique eight-column interlocking device,flip-style switch cover,touch-type handle. |
Controller and Memory Storage Function | Uses a microcomputer intelligent control system.It has 6 default operating modes,82 customizable modes,and 88 memory storage programs. |
Safety Devices | Door lock interlock device;cover closure check;dry burn protection;water level sensor monitoring;electronic over pressure protection;mechanical safety valve;over temperature and heating protection;over current,short circuit protection;leakage protection;cooling lock;burn injury safety protection;automatic fault detection system. |
Accessories | Condensate collection bottle,water separation plate,drainage hose,stainless steel basket. |
Optional Accessories | Printer,cooling fan,item temperature sensor,USB output. |
Series Description | A series(standard type )/ B series(auto water replenishment type) / C series(auto drying type) |